Will gc be able to detect glycerin in samples of beer? Should I use a capillary or packed column?
![]()
![]()
![]()
![]()
By Molever on Thursday, March 9, 2000 - 08:27 am:
We do glycerine and similar routinely here by capillary GC. Make sample 2% in DMF, mix 500 ul with 250 ul BSTFA, methyl silicone capillary column, no problem at all. I published this for glycerine in soap products in 1987 (I believe) in the JAOCS. External standard quantitation is fine, or use butanediol as internal standard.
![]()
![]()
![]()
![]()
By Anonymous on Thursday, March 9, 2000 - 09:07 am:
What is the JAOCS?
![]()
![]()
![]()
![]()
By Molever on Thursday, March 9, 2000 - 10:32 am:
Journal of the American Oil Chemists Society, Vol. 64, No. 9 (September 1987), pp. 1356-1357.
![]()
![]()
![]()
![]()
By Anonymous on Monday, March 13, 2000 - 04:23 am:
Cheers Molever.