High HPLC column pressure

Chromatography Forum: LC Archives: High HPLC column pressure
Top of pagePrevious messageNext messageBottom of pageLink to this message  By Anonymous on Thursday, June 8, 2000 - 03:36 pm:

is there any way other than back flush the column to decrease the pressure of a HPLC column?


Top of pagePrevious messageNext messageBottom of pageLink to this message  By H W Mueller on Thursday, June 8, 2000 - 11:25 pm:

If backflushing doesn´t work, replace the entrance frit (careful!). If proteins, etc., are involved wash until the pressure drops to an acceptable level with 3% Li-dodecylsulfate (aqu.) containing 0.01-0.02M 1,4-dithiothreitol. Incidentally, there are as many methods as chromatographers.


Top of pagePrevious messageNext messageBottom of pageLink to this message  By guido on Friday, September 1, 2000 - 01:20 am:

USE A PRECOLUMN!! I change my precolumn on a regular schedule(every 300 to 500 injections) and i seldom have any trouble with my analytical column. I inject thousands of plasma extracts and peak shape, separation and backpressure are as good as with a new column.


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Uwe Neue on Saturday, September 2, 2000 - 09:09 am:

Applause !!! This is exactly what we are doing and we got column lifetimes exceeding 10,000 injections.....


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Anonymous on Friday, October 27, 2000 - 07:26 am:

If the column is not very expensive, use a new column. The results will be better.


Top of pagePrevious messageNext messageBottom of pageLink to this message  By PGPG on Saturday, October 28, 2000 - 06:38 pm:

Flush with hot water