1.
Selectivity and/or retention time changes
Retention time
changes can be caused by a wide range of chemical and/or physical
problems with the chromatography.
For isocratic
separations, the first thing to look at is whether the retention
times of all the peaks (including the dead time of the system) changed
by the same percentage. If so, this is diagnostic of an
incorrect flow rate. We recommend the following steps:
- Confirm that
your
pump flow rate is correctly set.
- If the
retention
times are longer than specified, check the system carefully for
evidence of leaks.
- If the
retention
times are longer than specified, check the system for evidence of
bubble problems (inadequate degassing) or check valve failure; these
will often be accompanied by excessive fluctuations in the baseline or
the system pressure.
For isocratic
separations, if the dead time meets specification and/or if the
retention times of different peaks in the separation have changed by
different percentages, then a change in system chemistry should be
suspected. We recommend the following steps:
- Check
that the column temperature is correct (if the column
compartment temperature setting is correct, use a thermometer to
confirm the actual temperature).
- Check to
confirm
that the mobile phase was correctly prepared (especially if the problem
coincided with a new batch of mobile phase!).
- Check to make
sure that the correct type of column is being used.
- If the column
is
nearing the end of its useful lifetime, replace the column.
- If there is
a trend to the change (i.e., if retention or selectivity changes in the
same fashion with each injection, check to see if the change
correlates with the number of injections or with the volume of mobile
phase pumped (carry out a series of injections with different delay times
between injections, then plot retention time or alpha as a function of
time and as a function of number of injections).
- If the trend
correlates with number of injections, you should suspect that some
sample component is building up on the column.
- If the trend
correlates with time (i.e., with volume of solvent pumped), you should
suspect that some mobile phase component is building up on the column
or that the column is deteriorating under your separation conditions.
For gradient
separations, if the retention times of all the peaks have changed
by the same amount, but the dead time of the system is still
within specifications, a change in dwell volume is a strong possibility.
Check to see whether the instrument plumbing has been changed recently
(for example, substitution of a larger or smaller injection loop).
For gradient
separations, a change in selectivity can be caused by a number of
factors:
- Flow rate
differences can affect selectivity in gradient separations (unlike
isocratic separations).
- A change in
instrument dwell volume can affect selectivity,
particularly for early-eluting peaks. Check your
dwell volume to confirm that it meets specifications.
- A change in
operating temperature of the column is always a possibility. Check that
the column temperature is correct (if the column compartment
temperature setting is correct, use a thermometer to confirm the actual
temperature).
- Differences in
column equilibration at the beginning of the run can affect retention
and selectivity. Check to make sure that the column equilibration time
is exactly the same before each run. Alternatively, confirm that the
equilibration time has been sufficient to completely equilibrate the
column.
- A
malfunction in the pumping system or proportioning valve may result in retention
and/or selectivity changes. This is very likely if the problems are
intermittent.
- Check to
confirm that the mobile phase was correctly prepared (especially if the
problem coincided with a new batch of mobile phase!).
- Check to make
sure that the correct type of column is being used.
Ref: LC-GC, 15(9)
826 (1997)
Ref: LC-GC, 17(10) 908 (1999)
©
2006 LC Resources, Inc. all rights reserved