Contents     Index   

3. Abnormally wide peaks (abnormally low plate count)
Peak broadening in a chromatographic separation can be characterized by one of a number of peak width measurements or by the plate number, with wider peaks (for a given retention) having a lower plate number.

If all the peaks in the chromatogram are abnormally broad, then the following possibilities should be checked:

1. For isocratic separations only, especially if retention times are too long:
A. Is the pump flow rate is correctly set?
B. Check the system carefully for evidence of leaks.
C. Check the system for evidence of bubble problems or check valve failure; these often will be accompanied by excessive fluctuations in the baseline or the system pressure.

2. For both isocratic and gradient separations:
A. Is the temperature of the column compartment correctly set?
B. If the column compartment temperature is correctly set, use a thermometer to confirm the actual temperature.
C. Check whether the dilution solvent is stronger than the mobile phase.

3. For gradient separations only: check for excessive extra-column volume

4. For all separations, chemical contamination or attack on the stationary phase is always a possibility. Replacing the column should fix the problem; in most cases, this is the most cost-effective solution.

If only some of the peaks in the chromatogram are abnormally broad, while other peaks meet the peak width specifications, you should check the following possibilities:

1. Sample overload (especially if the broad peaks have a significantly larger area than the narrow peaks).

2. Incorrectly prepared mobile phase (pay special attention to buffer concentration and pH).

3. A partially-resolved interference.

4. Sample degradation.


Ref: LC-GC 17(9) 824 (1999)



© 2006 LC Resources, Inc. all rights reserved